• Run the agitator during operation.
• In case of gas pressure during operation, pressure up the ves-
sel for calibration, or calculate the absorption of the gas den-
sity.
• Cooling and heating jackets must be filled for calibration. In
order to get the density of the cooling/heating fluid, it should
be at the same temperature as during operation.