7 Error Handling Volume 6
17 Error RTCbackup / Battery Hardware timer does not have the
current date!
If the device has been without
power supply for a longer period of
time, then the buffer capacitor may
be empty.
Check date and update it, if neces-
sary. Supply device with power for
24 hours to charge the capacitor
again.
If the error message occurred while
the Uni-Probe has been in operation
for several hours already, then the
hardware timer is faulty.
You can continue working with the
device anyway, as long as it will be
continuously supplied with power.
Otherwise, please keep in mind that
no decay compensation is carried
out as long as the Uni-Probe is with-
out power supply. As soon as possi-
ble, you have to replace the digital
board to ensure that the decay
compensation will also work when
the device is switched off.
18 Error RTC Hardware timer faulty Replacing the digital board
19 Error RS-485 Communic. Communication not possible with
indicated Auxiliary Unit.
Check wiring. If necessary, replace
Main Unit and/or Auxiliary Unit
20 Caution cps < Min Count rate below adjusted mini-
mum
Check input value for minimum
count rate. If necessary, the func-
tion can be disabled by entering 0.
21 Caution cps > Max Count rate above adjusted maxi-
mum
Check input value for maximum
count rate.
If necessary, the function can be
disabled by entering 0.
22 Warning Rad.Interference Interference radiation detected,
possibly due to welding check.
Wait for set waiting time.
The system automatically switches
to the Run mode.
Error
number
Error message Probable cause Potential solution