4 Repair, Maintenance and Service Volume 2
Replacement of the multiplier requires a new setting of detector
code and HV parameters. Please see Volume 2, "1.6 Detector
Codes", page 2–133, and Volume 3, "2.38 HV SETTINGS", page 3–
334.
Check the level indication empty and full. If you detect any
deviations, please carry out a new calibration, see Volume 3,
"4 Calibration", on page 3–341.
Separate User’s Manuals are available for user interfaces such
as SIMATIC PDM or FOUNDATION
™
Fieldbus.
This completes the replacement of the crystal-multiplier
assembly.