Volume 3 4 Calibration
3
Select CALIBRATION I/O.
One- and two-point calibration
Select NO. OF CAL. POINTS and enter the number of calibration
points.
One-point calibration: 1 calibration point
Two-point calibration: 2 calibration points
Save the changes with
SEND.
Select
1-2 POINT CURVE.
Multi-point calibration
Select NO. OF CAL. POINTS and enter the number of calibration
points.
Multi-point calibration: 3 -10 calibration points
Save the changes with
SEND.
Select
MULTI POINT CURVE.
Tip
The number of calibration points can be changed later in order to
record further calibration points.
This completes the setting of the operating mode for the cal-
ibration.