Uni-Probe LB 490
BERTHOLD TECHNOLOGIES GmbH & Co. KG
5 – 501
Volume 5 5 Calibration with SIMATIC PDM
5
Confirm averaging time with OK.
Click OK to start reading-in the pulse rate.
The count rate reading-in process has started. If the process
has been started with 0, it will take 30s until the count rate is
read in.
Confirm the pulse rate with
OK.
Confirm the dialog with
OK.
This completes the empty calibration.
5.3.3 Full Calibration
Select DEVICE CALIBRATION ADJUST UP. CAL.PT.