6 – 607
Volume 6 4 Calibration with SIMATIC PDM
6
Select the CALIBRATION OPTIONS tab.
Enter the number of calibration points in the
NO. OF CAL. PTS
box.
One-point calibration: 1 calibration point
Two-point calibration: 2 calibration points
Multi-point calibration: 3 -10 calibration points
to upload the settings to the Uni-Probe, Click
TRANSFER.
Click CLOSE to exit the dialog box.
The number of calibration points can be changed later in order to
record further calibration points.
This completes the entry for the operating mode for calibra-
tion.