Volume 7 4 Calibration with FOUNDATION
Fieldbus
7
4.2.1 Setting the Type of Linearization for Calibra-
tion
IMPORTANT
The function Exponential is used only in special cases. For informa-
tion on the calibration modes, please refer to chapter 6.3.
The type of linearization can only be defined for one-point and two-
point calibrations. See also section 6.3.
Go to the function block
TRANSDUCER.
At LE_LIN_TYPE enter the desired type of linearization: Linear
or Exponential.
Send the entry to the Uni-Probe.
This completes the entry of the type of linearization.
Parameters used
LE_LIN_TYPE