5 Functional Processes Volume 7
Select Start plateau measurement at
LE_START_STOP_PLATEAU and send the entries to the Uni-
Probe to start the measurement.
•
LE_PLATEAU_STATUS shows the progress of the plateau mea-
surement.
LE_PLATEAU_HIGH_VOLTAGE shows the currently
used high voltage,
LE_PLATEAU_PULS_RATE shows the currently
is measured count rate. As soon as the range defined above has
been covered,
LE_PLATEAU_STATUS changes to Read In Off.
You may view the individual data points via
LE_PLATEAU_DATA_ENTRY and LE_PLATEAU_DATA: Enter the
desired data point, e.g. 3 for the third value, and send the entries
to the Uni-Probe. The record LE_PLATEAU_DATA includes in
LE_PLATEAU_DATA_HIGH_VOLTAGE the HV voltage and in
LE_PLATEAU_DATA_SENSOR_VALUE the measured (averaged)
pulse rate.
This completes the plateau measurement.