6 Explanations Volume 7
• Run the agitator during operation.
• In case of gas pressure during operation pressure up the vessel
for calibration, or calculate the absorption of the gas density.
• Cooling and heating jackets must be filled for calibration. In
order to get the density of the cooling/heating fluid, it should
be at the same temperature as during operation.