Chapter 1 Introduction
System Components
The ZE5 Cell Analyzer system includes the following components:
n ZE5 Cell Analyzer Instrument (one of the following)
ZE5 Cell Analyzer, 5 laser 27 color
(355/405/488/561/640)
ZE5 Cell Analyzer, 3 laser 17 color
(405/488/640)
ZE5 Cell Analyzer, 4 laser 24 color
(405/488/561/640)
ZE5 Cell Analyzer, 3 laser 17 color
(488/561/640)
n ZE5 Computer System
ZE5 Cell Analyzer Computer with Network
Adaptor
ZE5 Cell Analyzer Wireless Keyboard and
Mouse
ZE5 Cell Analyzer Computer Monitor, 29 in,
2560 x 1080
n ZE5 Accessory Kit
The Accessory Kit includes consumables, documentation, and accessories for initial instrument
startup.
For information on consumables and accessories, see Appendix E, Ordering Information.
Installation Requirements
The ZE5 Cell Analyzer should be installed by a Bio-Rad service engineer to ensure proper instrument
operation and calibration. If any items are missing or damaged, contact your local Bio-Rad office.
Choose an appropriate site for ZE5 Cell Analyzer installation, such as a sturdy bench or tabletop,
away from any other instruments that might interfere electrically, or mechanically by causing vibration.
The bench or tabletop must accommodate 275 lb (125 kg), the approximate weight of the instrument,
computer, and monitor. The area should be free of excessive dust or moisture. Do not place the
instrument near air conditioning or heating vents, strong electrical fields, magnetic fields, or equipment
that can produce vibrations, for example, bench centrifuges.
Table 2 on page 21 lists the benchspace required to install the ZE5 Cell Analyzer with and without the
computer and monitor.
20 | ZE5 Cell Analyzer and Everest Software

Table of Contents