Chapter 12 Maintenance
5. Approximately 50 min into the procedure, the software prompts you to replace both sheath bottles
with bottles containing 2.5 L of DI water, and replace the additive and cleaner bottles with bottles
containing 200 ml of DI water.
6. After replacing the bottles, click Continue.
7. Approximately 190 min into the procedure, the system prompts you to replace both sheath bottles
with bottles filled to the Fill Line with sheath fluid, and replace the additive and cleaner bottles
with the default (original) bottles filled to the Fill Line with the appropriate solutions.
332 | ZE5 Cell Analyzer and Everest Software