EasyManua.ls Logo

BIO RAD ZE5 - Page 343

BIO RAD ZE5
398 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Deleting Acquired Data Files
n At least one folder labeled with a global unique identifier (GUID)
n The experiment’s run list (.rlst) file (5Color_2.rlst in this example)
n A run list status (.rlstatus) file for each GUID folder
n The instrument.definition file
The GUID folder contains the FCS data that you recently acquired for the session.
7. If you have more than one GUID folder, open each folder and locate those that contain a folder
labeled fcs.
8. Delete the GUID folder or folders that contain an fcs folder.
9. Log in to Everest Software.
Your experiment appears in the Recent Experiments panel. You can edit and run the experiment,
and acquire data again.
Note: Because you deleted the acquired data, Everest Software does not display data for the
experiment in Analysis mode, for example:
User Guide | 341

Table of Contents