EasyManua.ls Logo

Bruker IFS 125M - Page 153

Bruker IFS 125M
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
153
Bruker Optik GmbH IFS 125M User Manual
Web Interface 8
8.3.2.3 Scanner diagnostics page
The Scanner line includes the prominent positions of the scanner on its scan travel.
Front: Position of the front light switch. This position is by definition always 0.
Peak: Peak position relative to the front position in half laser periods.
Center: Position half between front and back
Back: A fixed position at 4,000,000 half laser periods
Laser A Ampl. [mV] and Laser B Ampl. [mV]: All laser amplitude values should be
>1800
mV and <18000 mV. If they are <1800 mV, the laser signal needs to be aligned. If
the values are >18000
mV, the laser beam needs to be attenuated.
Laser A Offs. [mV] and Laser B Offs. [mV]: Values smaller than -8500 mV indicate that
no laser light reaches the interferometer block. Cause: The laser is switched off or the
laser beam is out of alignment. A great difference between the Life max. value and the
Life min. value (>1000
mV) indicates that the laser beam is not properly aligned
(chapter
6.2).
i
As a rough rule of thumb, the scanner block temperature is a few °C above the ambient
temperature when thermally equilibrating the system. The ambient temperature range
for an operating spectrometer is specified between 18°C and 35°C. Consequently, a
scanner block temperature above 40°C indicates that the spectrometer is not operated
within the specified temperature range. There is no humidity sensor inside the spectrom
-
eter.
Definition
A
Status message
Table 8.9: Components of scanner diagnostics page
Figure 8.10: Scanner diagnostics page
A

Table of Contents

Related product manuals