EasyManuals Logo

Dräger Fabius User Manual

Dräger Fabius
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #145 background imageLoading...
Page #145 background image
Instructions for Use Fabius MRI SW 3.n 145
Configuration
Current settings
The current settings cannot be saved as default
settings.
To change the site defaults follow the instructions
on pages “Restore Site Defaults” on page 139 to
page 141.
Restore Factory Defaults
z Select and confirm the »Restore Factory
Defaults« label on the Standby Setup screen.
The Restore Factory Default setting window
appears on the screen.
z Select and confirm »Yes« or »No«. When
»Yes« is selected, the factory defaults are
restored and replace the current default set-
tings.
The factory defaults for the Fabius MRI are shown
in the table below:
Parameter Factory Default Settings
Volume
Control
P
MAX = 40
V
T = 600
Freq = 12
T
I:TE = 1:2.0
T
IP:TI = 10
PEEP = 0
Pressure
Control
PINSP = 15
Freq = 12
T
I:TE = 1:2.0
Insp Flow = 30
PEEP = 0
Pressure
Support
ΔPPS = 10
Freq Min = 3
Trigger = 2
Insp Flow = 30
PEEP = 0
541

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals