Perform calibration
Fig. 82: Software setting: ‘Calibration’
[Start calibration] :
1.
[Main menu] Press the button.
ð
The [Main menu] is displayed.
2. [Select calibration] .
ð
The screen for starting [calibration] is displayed.
3. [Press Start] .
ð
The screen changes to ‘Calibration step 1
.
Calibration step 1 runs automatically.
The ‘Calibration step 1’ screen is displayed.
4. Enter the quantity in litres (before the decimal point) using the scrollbar
.
5. Select the quantity in millilitres after the decimal point.
6. Enter the quantity in millilitres (after the decimal point) using the scrollbar
.
7.
[Click Next] .
ð
The screen changes to ‘Calibration step 2
.
The ‘Calibration step 2
runs automatically.
The screen page for calibration step 2 is called up.
8. Enter the quantity in litres (before the decimal point) using the scrollbar
.
9. Select the quantity in millilitres after the decimal point.
10. Enter the quantity in millilitres (after the decimal point) using the scrollbar
.
11.
[Click Next] .
ð
The setting is saved, and the screen switches back to the [operating display]
.
Start-up
138417102276 Rev. 5-02.2020