Planned maintenance check list
9
Document no. M1125633-06
NOTE: Avoid contact with the rubber paper roller. Be careful to limit the application of alcohol to
the thermal printhead and the window.
Reinstall the paper roll.
"
2.1.6 Modules with NIBP measurement
1. Check the NIBP filter. Replace the filter, if necessary.
Plug the module back into the frame.
"
2.2 Functional inspection
2.2.1 General
1. F-CU5(P): Connect all the parts together. Check that the cables and screws are tightened
properly. Connect the mains power cord to the N-AC Power Unit.
F-CU8: Connect the mains power cord to the F-CU8.
Check that the stand-by LED is lit.
2. Switch the monitor on.
Check that the monitor starts up properly, i.e. the alarm LEDs turn on shortly, normal
start-up sound is heard from the loudspeaker and the monitoring screen appears.
No error messages should appear on the screen.
3. Configure the screen for the parameters that are connected.
Check that all the connected modules are recognized, i.e. the required parameter
information is shown on the screen.
NOTE: InvBP waveforms are not shown without a parameter simulator.
If connected, the recorder should record two lines of start-up information.
Preset the measurement settings for those parameters that are connected, for example:
Others - EEG&EP - Montage - EEG Channels - 4
- Montage type - Bip
Others - EEG&EP - EEG Setup - Numeric 1 - MF
- Numeric 2 - Ampl.
Others - EEG&EP - Cycle - Cont.
- AEP Setup - AEP Channels - 2
- Responses -100
- Stim. Frequency - 1.1Hz
- Stim. Intensity - 90 dB
- Sweep length - 100 ms
- EP size - 1
or
OthersBIS – Scale - 100uV
Smoothing Rate 30s
BIS Setup –Automatic Check –ON