Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
56
Document no. M1137263-04
4.3 Interfacing Datex-Ohmeda Anesthesia Delivery Unit
It is possible to interface the Datex-Ohmeda Anesthesia Delivery Unit to the S/5 Anesthesia
Monitor via the UPI Board, B-UPI4 or UPINET Board, B-UPI4NET. The data link is bi-directional.
The parameters transferred to the Anesthesia Delivery Unit are summarized in Table 7 and the
events transferred to the S/5 Anesthesia Monitor are summarized in Table 8. The events are
transferred continuously once a minute or during the system check procedure and appear in
the trend pages or in the anesthesia record. The transfer of events requires record-keeping
software.
4.3.1 Interconnection
Use the ADU-AS/3 Monitor Cable.
4.3.2 Setting interfacing parameters on the S/5 Anesthesia Delivery Unit
1. Press the Setup key.
2. Select Install password 10.
3. Select Interfacing and choose S/5 AM.
4. Press the
Normal Screen key.
4.3.3 Setting interfacing parameters on the S/5 Anesthesia Monitor
No settings are required on the S/5 Anesthesia Monitor.
Table 7 Parameters transferred from S/5 Anesthesia Monitor to S/5
Anesthesia Delivery Unit
Waveforms Numerics
CO
2
Et & Fi CO
2
%
Et & Fi O
2
%
O
2
% difference
(I-E)
Et & Fi N
2
O%
Et & Fi AA%
Vol Exp MV
Vol Exp TV