Displays and Controller Boards
17
Document no. M1137272-02
10. Functioning after safety check
Check that the LCD display functions normally after the performed electrical safety check.
"
11. Final cleaning
Clean the LCD display with suitable detergent.
"
Fill in all necessary documents.
3.2.2 15” Flat Panel Display, D-FPD15
Tools needed
Central Unit, B-CPU6 or B-DISPX and K-ANEB/K-ICUB Command Bar.
Visual inspection
Turn off the monitor and display.
1. Video cable
Check that the video cable is intact.
Check that the video cable is properly connected and locked both to the monitor
board and to the display with the locking screws.
Check that the power cord is intact and properly in place.
"
2. External parts
Check external parts:
the outer cover is intact
the display screen is intact
the desk stand is intact and tightened properly
the keypad holder is intact and tightened properly
"
Functional inspection
Turn on the monitor and press the Control dial to turn on the LCD display.
3. Resolution
Wait until the normal monitoring screen appears on the LCD display.
Check that the picture on the LCD display screen is clear and stable.
NOTE: Check that the display screen uses the XGA resolution.
Monitor Setup - Install/Service (password 16-4-34) - Installation - Display Setup