Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
60
Document no. M1137263-04
4.5 Interfacing printer
It is possible to interface a laser printer (either serial or parallel) to the S/5 Anesthesia Monitor
and S/5 Critical Care Monitor via the UPINET Board, B-UPI4NET. The printer must be PCL5 or
PCL6 compatible and it should contain at least 2 MB of memory.
4.5.1 Interconnection
Use a commercial serial printer interface cable or the UPI4 - Printer parallel interface cable
713701 or a standard parallel port printer cable for connecting the printer to the monitor.
1. Make sure that the power on both units is turned off.
2. Connect the cable to the corresponding connector on the B-UPI4NET.
3. Connect the cable to the corresponding connector on the printer.
WARNING Always make sure that the combination complies with the international
safety standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
WARNING Connecting the power supply cord of the printer to the wall power outlet
may cause the printer leakage current to exceed the limit specified for
medical equipment. A printer must be supplied from an additional
transformer providing at least basic isolation (isolating or separating
transformer).
4.5.2 Setting interfacing parameters on the printer
NOTE: For more information about printer settings, please refer to the documentation provided
with the printer.
4.5.3 Setting interfacing parameters on the S/5 Anesthesia Monitor or
S/5 Critical Care Monitor
Select the appropriate printer connection in the Record/Print menu.
4.6 Interfacing computer
It is possible to interface a computer to the S/5 Anesthesia Monitor and to the Critical Care
Monitor.
Contact your authorized GE Healthcare distributor for further advice on computer interface.
WARNING Always make sure that the combination complies with the international
safety standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
WARNING Connecting the power supply cord of the computer to the wall power outlet
may cause the computer leakage current to exceed the limit specified for
medical equipment. A computer must be supplied from an additional
transformer providing at least basic isolation (isolating or separating
transformer).