5-Module Frame, F-CU5, F-CU5P
21
Document no. M1125636-05
17. Optional display controller board B-DISPX jumpers/DIP switches
If available, disconnect the display controller board(s) and check that the address and
interruption jumpers/ DIP switches have been set correctly. The instructions for setting
the jumpers can be found in the”Part I, System Installation” slot of the Technical
Reference Manual,
Leave the board(s) disconnected.
"
F-CU5 Power Unit, N-AC
18. Power unit frame
Check that the power unit frame is intact.
"
19. Power output connectors
Check that the power output connector is clean and intact.
"
20. Power inlet connector
Check that the power inlet connector is clean and intact.
"
21. Equipotential tap and rear panel screws
Check that the equipotential tap and all the rear panel screws are tightened properly.
"
3.2.4 General inspection
22. Communication lines
After any service repair for the Module mother board, measure resistance from the
following Module mother board’s connector pins against the ground:
Check that the resistance on each of the pins is higher than 10 k. If not, replace the
Module mother board.
Reconnect the PC boards and secure them to the Central Processor Unit with screws.
Attach the F-CPU Power cable to the N-AC and F-CPU.
Attach the F-CPU Signal cable to the F-CPU and F-CU5(P).
Module mother board connector Pin 1 +Reset RS485
Pin 5 -Data RS485
Pin 6 +Data RS485
Pin 8 -Reset RS485
Pin 13 Ground
13
1
14
25

Table of Contents