Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
4
Document no. M1137266-02
1.1 Display Controller
Contro
ll
er 1
d
eta
il
s o
f
t
h
e 1
s
t
di
sp
l
ay contro
ll
er.
Controller 2 details of the 2
nd
display controller.
Controller 3 details of the 3
rd
display controller.
Previous Menu a selection to return to the previous
menu.
Details of the display controller boards
NOTE: The values are valid only for B-DISPX and newer
display controller boards.
Board type: This is the type of the display controller
board in the monitor.
Serial number: Serial number of the display controller
board in use.
PLD version: Program version of the programmable
logic device.
SW date: Manufacturing date of the display controller
software.
SW version: Version of the display controller software.
NOTE: The display controller software is a part of the
monitor software, and it is downloaded to the display
controller board memory from the CPU board when
necessary.
Resolution: This shows the output resolution that is
selected through the monitor software menus.
Forced resolution: Forced resolution is the output
resolution selected by the display controller board’s
dip switches or jumpers.
NOTE: The forced resolution settings override the
resolution settings selected through monitor menus.
The forced resolution (i.e. resolution dip switches)
must be set to AUTO to allow the software-selected
resolutions to be valid.