Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
34
Document no. M1137263-04
3.5 Displays
Figure 16 Display options
3.5.1 Main displays
The main display options are integrated with an S/5 Command Bar and a ComWheel:
15” Flat Panel Display, D-FPD15
S/5 LCD Display, D-LCC12A
NOTE: Only one LCD display (D-LCC12) that uses B-DISPX as the power supply can be used in a
system where one F-CU5(P) 5-Module Frame is connected to Central Unit F-CPU.
NOTE: An LCD display that uses B-DISPX as the power supply (D-LCC12) cannot be used in a
system where two F-CU5(P) Central Units are connected to one F-CPU.
3.5.2 Secondary displays
19” medical grade display
However, the secondary display can also be either of the main displays listed above.
3.5.3 3
rd
display
The 3
rd
display can be installed together with the Display Controller Board, B-DISPX. All S/5
main softwares support the use of the 3
rd
display.
NOTE: Due to increased power consumption only two LCD displays should take power via the
Central Unit simultaneously.
3.5.4 Display installation
NOTE: Please pay attention to any additional instructions that may accompany the displays
and / or their installation equipment. The instructions described below in this manual may not
contain all the necessary information.
NOTE: When connecting a display with a DVI-D type digital video cable, some adjustments are
not needed and may not be available in the display’s on-screen adjustment menu (OSD).
WARNING A secondary display must be supplied from an additional transformer
providing at least basic isolation (isolating or separating transformer).
D-LCC12A
D-FPD15
19’’ medical grade display