Service Menu
9
Document no. M1137266-02
1.4.2 Communication
Interface status indicates the status of the data link
between the monitor and memory module. If the
memory module is properly attached, the status
should always be on ACTIVE. If the status blinks
between ACTIVE and CLOSED, a communications
error has occurred: remove the module briefly, and
insert it back to the monitor frame to check if the
error disappears.
Message types indicates the type of data packets
that have been sent (Tx) and received (Rx) since last
monitor start. Data types are listed on the lines below
Message types text.
Data types:
Record K indicates the communication between the Monitor and Record Keeper.
File Operation indicates the operations of Patient data.
Service indicates the Memory Module operations.
Modes indicates the User Mode operations.
Module status indicates the number of sent/received data packets that relate to the memory module
status.
Packets total indicates the total amount of data packets that have been sent/received since the last
monitor start.
Bytes total indicates the total amount of data bytes that have been sent/received since the last monitor
start.
The last four lines indicate transmission errors:
Timeouts indicates the number of timeouts that have occurred in memory module data transmission
since the last monitor start.
Chksum err indicates the number of checksum errors in data packets from memory module since the
last monitor start.
Length err indicates the number of data packets with erroneous length from the memory module since
the last monitor start.
Duplicated indicates the number of duplicate data packets from the memory module since the last
monitor start.