Datex-Ohmeda S/5
TM
Anesthesia and Critical Care Monitors
6
Document no. M1125639-03
3. ComWheel cover
Check that the ComWheel cover is intact and attached properly.
"
4. ON/STBY switch
Check that the ON/STBY switch is intact and attached properly. Check that the ON/STBY
switch changes its state firmly when turning it back and forth. Leave the switch into STBY
position.
"
5. Stand-by LED
Reconnect and lock the command bar cable to the B-CPU6 board.
Check that the green stand-by LED is lit up (the Central Unit power cord is connected to
the mains).
"
3.2.3 Functional inspection
•
Turn the monitor on and enter the service menu:
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8)
6. Keyboard version
Record the information regarding keyboard software.
"
7. Alarm LEDs
Select the menu KEYBOARD with the ComWheel.
Highlight the text UPPER LED. Check that the red alarm LED is turning on and off when
pressing the ComWheel. Check also the yellow alarm LED by selecting LOWER LED from
the menu.
"
8. Membrane keys
Press the keys one by one. Check that each key generates a sound from the loudspeaker
and the corresponding text in the menu changes from yellow to red.
"
9. ComWheel
Turn the ComWheel clockwise and counterclockwise and check that each step generates
a sound and the corresponding values at the bottom of the menu increase.
Select DUMMY PRESS. Press the ComWheel and check that the press generates a sound
and the corresponding value in the menu increases.
"