Displays and Controller Boards
19
Document no. M1137272-02
3.2.3 19” medical grade display
Tools needed
Central Unit, B-DISPX or B-CPU6 and Keyboard K-ANEB/K-ICUB.
Visual inspection
Turn off the monitor and display.
1. Interface cable
Disconnect and check the LCD display interface cable:
The monitor side connector:
− the connector pins are clean and straight and at about the same height
− the locking screws are intact
The display side connector:
− the screw(s) on the casing is tightened properly
− the connector pins are clean and intact
− the locking screws/claws are intact
Check also that the cable itself is intact.
"
2. Check external parts
− the outer cover is intact
− the display screen is intact
− the block screws for the cable are in place and are tightened properly (if installed)
− the block screw threads are intact (if installed)
"
Functional inspection
3. Power cord
Connect and lock the LCD interface cable to the LCD display and to the board in the
monitor.
Check that the display power cord is intact and properly in place.
• Turn on the monitor and press the Control dial to turn on the LCD display.
"
4. LCD display picture
Wait until the normal monitoring screen appears onto the LCD display.
Check that the picture on the LCD display screen is clear and stable.
NOTE: Check that the display screen uses the SXGA resolution.
NOTE: The SXGA resolution must be set every time after Factory Reset.
Monitor Setup - Install/Service (password 16-4-34) - Installation - Display setup