Functional check
71
Document no. M1137263-04
5.3.12 Masimo Compatible Saturation module, E-MASIMO
1. Connect a Masimo SpO2 finger probe to the module. Check that the message ‘Check
probe’ is displayed on the screen within 30 seconds.
2. Attach the SpO2 probe on your finger. Check that a reading of 95-100 and a proper SpO2
waveform appear.
"
5.3.13 Nellcor Compatible Saturation module, E-NSATX
1. Connect a Nellcor SpO2 finger probe to the module. Check that the message ‘Pulse
search ’ is shown and check that the message 'No probe' changes to ‘Check probe’ within
30 seconds.
2. Attach the SpO2 probe on your finger. Check that a reading of 95-100 and a proper SpO2
waveform appear.
"
5.3.14 BIS Module, E-BIS
1. Connect the BIS module to the monitor frame.
Check that ‘Cable off’ is displayed in the BIS waveform field.
2. Connect the DSC to the module.
Check that the ‘No sensor’ message appears in the waveform field.
3. Check the function of the front panel membrane keys.
4. Open the BIS Setup menu:
BIS - BIS Setup and check that Test DSC shows PASS.
NOTE: If the sensor is not connected, the
Check Sensor key is inoperative.
"
5.3.15 Entropy Module, E-ENTROPY
1. Check the function of the front panel membrane keys.
2. Connect the Entropy sensor cable and Entropy simulator to the module. Check that
‘Checking sensor’ text and an image appear in the waveform numeric field. Wait for a
while and check that all sensors show PASS.
3. Check that the EntrEEG waveform and RE and SE values appear on the monitor screen.
"
5.3.16 Memory Module, E-MEM
1. Insert a Data card or a Menu card to the slot.
Check that the corresponding symbol appears on the monitor screen.
"