Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
38
Document no. M1137266-02
4.7.7 NIBP Watchdog
How to test Adult watchdog time
Select Test ADULT and press the ComWheel. Watchdog interval starts counting up seconds
and keeps on counting as long as the +15 Vdirty is on. The time should be 120 to 140 seconds.
How to test Infant watchdog time
Select Test INFANT and press the ComWheel. Watchdog interval starts counting up seconds
and keeps on counting as long as the +15 Vdirty is on. The time should be 60 to 70 seconds.
These menu items have only effect on the M-NIBP
modules. They do not effect other NIBP modules.
Test ADULT: is to test watchdog timer in adult mode
(120 to 140 seconds).
Test INFANT: is to test watchdog timer in infant
mode (about 60 to 70 seconds).
Stop Test: is for stopping the test.
Watchdog Data field
For information on general items Pressure, Zero,
Protect handle, Calibr. prot., +15 V power, AD0 to
AD7 as well as Timeouts etc., see section “4.7 NIBP
Module”.
Watchdog Interval: shows the time the +15 Vdirty
stays on during the test.