Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
62
Document no. M1137263-04
If the output signals are used simultaneously with the coding element, the B-UPINET Y-cable,
order number 889308, is recommended.
4.7.3 Analog outputs
Direct ECG B-UPI4NET -02 (X7 pin 8)
Delay (max.): <15 ms
Gain ECG (out)/ECG (in): 1 V/mV ±20%
Pacer: 5 V and 2 ms pulse
Output range: ±4 V
Noise: 50 mVpp max.
The signal requires input impedance of 100 kΩ.
NOTE: The ECG signal is based on the ECG measurement of the Hemodynamic Modules,
E-PRESTN (all variations) and E-PSM / E-PSMP. The ECG signal from the modules is channel 1
(ECG1). The channels 2 and 3 (ECG2, ECG3) are not transmitted. Make sure that the signal in
channel 1 (ECG1) is good enough (extensive QRS for IABP synchronization). For further
information, please contact your authorized GE Healthcare distributor.
Pressure out B-UPI4NET -02 (X7 pin 7)
Invasive pressure signal: From pressure labelled ‘Art’
Delay (max.): <35ms
Gain signal (out) / Pressure (in): 10 mV/mmHg ±20%
Output range: -0.4 V to +3.2 V
Noise: 50 mVpp max.
The signal requires input impedance of 100 kΩ.
4.7.4 S/5 Pressure Temp Module, E-PT, output signals
The signal output connector on the Pressure Temp Module, E-PT, can be used to interface some
models of IABPs to the S/5 Anesthesia Monitor and S/5 Critical Care Monitor. The pin
assignments are illustrated in Table 13. Please contact your local distributor for more
information.
WARNING Always make sure that the combination complies with the international
safety standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
Table 13 Signal output connector pin assignments
Output signals
ECG out, ECG1 1V / 1mV
Pressure out, P3 1V / 100mmHg
ECG out, ECG1 1V / 1mV
Pin Signal
1
2
3
4
ECG out, ECG1
Pressure out, P3
ECG out, ECG1
Ground