Appendix A ElectroMagnetic Compatibility
B3(6)
Document no. M1137263-04
Table 3 Guidance and manufacturer’s declaration – electromagnetic
immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The S/5™ AM or CCM is intended for use in the electromagnetic environment specified below. The customer or
the user of the S/5™ AM or CCM should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance
level
Electromagnetic environment - guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the S/5™ AM
or CCM, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
3 Vrms
150 kHz to 80 MHz
1 Vrms
(1
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
3 V/m
80 MHz to 2.5 GHz
1 V/m
(2
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
d 1.2 P=
d 3.5 P=
d 1.2 P=
80 MHz to 800 MHz
d 2.3 P=
800 MHz to 2.5 GHz
d =3.5 P
80 MHz to 800 MHz
d =7.0 P
800 MHz to 2.5 GHz