Functional check
67
Document no. M1137263-04
Airway Gas -Spirometry Setup -Scaling -Indep.
-Paw Scale –20
-Flow Scale –15
or
Ventil.-Spirometry Setup -Scaling -Indep.
-Paw Scale –20
-Flow Scale –15
Others -Resp Setup -Size -1.0
- Resp Rate Source – AUTO
- Measurement – ON
- Detection Limit – AUTO
5.3.2 Display(s)
1. Check that the picture on the screen is correct. Readjust the picture with the adjustment
knobs or with the display menu keys, if necessary.
"
5.3.3 Keyboard(s)
1. Tests with all the connected keyboards:
− Press the
Monitor Setup key. Turn the ComWheel in both directions and check that
the cursor in the menu moves correspondingly. Select Normal Screen and check
that the menu disappears from the screen.
Tests with the Command Bar:
− Press the
Help key. Turn the ComWheel in both directions and check that the cursor
in the menu moves correspondingly. Select Normal Screen and check that the
menu disappears from the screen.
Check the rest of the menu keys by pressing them one by one.
Tests with the ARK Keyboard:
− Enter the Keyboard service menu
− Check functioning of the ComWheel.
− Press all keys, except
Modify and Print. Check that each key produces a sound
from the loudspeaker, or the Message count value in the service menu increases.
− Press the
Modify and Print keys and check that the corresponding menus appear
on the monitor screen.
Tests with the Remote Controller:
− Enter the Keyboard service menu.
− Check the function of the ComWheel.
− Press all keys. Check that each key produces a sound from the loudspeaker, or the
Message count value in the service menu increases.
"