1.5 Power Supply
The menu shows the voltages and temperature
measured by the UPI4(NET) board. The measurement
starts about 100 ms after a start-up. The values in
the column under Mean are the mean values of the
last one second, the Min column shows the
minimum mean value, and the Max column the
maximum mean value of the voltages and
temperature measured during the current power
ON.
The voltages should meet the following ranges:
+15V 14.20….15.60
–15V -14.00...-15.50 V
+15VD 14.10...15.60 V
+5V 4.70...5.40 V
Temp (°C) value corresponds with the Central Unit
internal temperature measured at the location of the
UPI4(NET) board.
The numbers on this page are only directive and not
absolute values.