System installation
37
Document no. M1137263-04
Mounting to Display arm
Mount the LCD Display, D-LCC12A to a pedestal, wall or ceiling mount, or to a display arm with
appropriate parts. Connect the video cable to the display and tighten the video cable finger
screws carefully.
CAUTION To prevent any liquid from entering the LCD display casing, do not tilt the LCD
display more than 45 degrees backward or 20 degrees forward.
Connection to Central Unit
1. Make sure that the power to the Central Unit is turned off.
2. Connect the video cable to the X2 connector on the CPU board, B-CPU6. Tighten the
finger screws.
3.5.7 19” medical grade display
Environmental conditions
WARNING The environmental conditions of the 19" Medical Grade Display differ from
the environmental conditions of the S/5 Anesthesia Monitor and Critical
Care Monitor systems.
Atmospheric pressure lower limit
Operating: 697 mbar
The difference in the operating atmosphere pressure lower limit is only in altitudes over 3000 m
(9800 ft). Above this altitude, use another specified display.
Mounting
When using the display with the desk stand option, it must be placed on a flat horizontal
surface. The 19” medical grade display is provided with a desk stand as default with the GCX
adaptor included. Standard VESA mounting solutions can be used with this display.
In order to mount the display with AM and CCM monitors, please see for example the Supplies
and Accessories catalog.
Connection to the Central Unit
Medical grade display requires B-DISPX display controller board.
1. Make sure that power to the Central Unit is turned off.
2. Remove the connector cover: Press “fixed lever” downward on the back of the display.
Then slip the cover to the direction of the arrow and remove the cover.