Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
36
Document no. M1137263-04
Install/Service (password 16-4-34) – Installation - Display Setup
NOTE: After changing the resolution, the monitor must be restarted before the new setting
becomes effective. If the resolution setting is not correct after the monitor restart, check that
the resolution jumpers or dip switches on the display controller board are positioned correctly,
see chapter 3.6. Display controller boards.
2. Brightness and contrast adjustments
Press the Control dial while the LCD display is on to activate the OSD menu
(on-screen-display).
NOTE: Pressing the Control dial for more than 2 seconds turns the power off.
Perform the contrast and brightness adjustments if needed. Please refer to the display’s
User’s Manual.
3. Color temperature adjustments
Perform the color adjustments if needed. Please refer to the display’s User’s Manual.
4. Auto Setup
Perform Auto Setup to automatically adjust the screen size, position, phase and clock, if
needed.
This adjustment is available only with analog input (VGA cable). This function is not
available for digital input (DVI).
NOTE: Position adjustments, clock and phase can be adjusted also manually. Please refer to the
display’s User’s Manual.
Exit the OSD menu and go back to the normal screen.
3.5.6 12” LCD Display, D-LCC12A
(1) 12” LCD display, D-LCC12A
(2) Connector for the anesthesia record keeping solution keyboard or Remote Controller
K-REMCO
(3) ON/STBY switch of the monitor
(4) ComWheel
(5) Command bar keys
(6) Display Arm