Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
2
Document no. M1125635-05
1Specifications
1.1 8-Module Frame, F-CU8
Dimensions, D × W × H382 × 315 × 128 mm / 15.0 × 12.4 × 5.0 in
Weight 9.1 kg / 20 lb.
1.1.1 Electrical requirements
Rated voltages and and frequencies 100 V, 50/60 Hz
110...120 V, 50/60 Hz
220... 240 V, 50/60 Hz
Allowed voltage fluctuations 100 V ±10%
110 V -10% to 120 V +10%
220 V -10% to 240 V +10%
Maximum power consumption
F-CU8 Rev.12 160 VA
Safety class Class I
Grounding Hospital grade
Interruptibility Data memory and alarm settings are saved during power
failures up to 15 minutes
1.1.2 Environmental requirements
Operating temperature 10...35 °C / 50...95 °F
Storage temperature -10...+50 °C / 14...122 °F
Relative humidity 10...90% non-condensing
(in airway 0...100% condensing)
Atmospheric pressure 660...1060 mbar (500 to 800 mmHg)