4.1 Troubleshooting for the Ethernet section
Symptom at the monitor end Problem at Explanation/What to do
Monitor does not connect to the
network.
Monitor connects to the network,
but disconnects unexpectedly
(‘Network connection down’
message on the monitor screen).
Patch panel Patch cable not connected to HUB or to
panel.
Patch cable Patch cable or connector defective.
HUB not connected to power supply.
HUB port closed due to physical layer
problems.
HUB port temporarily closed and
reopened due to physical layer problems.
HUBs not properly connected to each
other.
Monitor-Network cable Cable not properly connected to the wall
plate or to the monitor.
Monitor-Network cable Cable or connector defective.
B-CPU6 board The B-CPU6 board is defective. The board
cannot be used. See network service page
for details.
Identification plug on
B-UPI4NET board
There is no identification plug attached to
the monitor.
The identification plug is defective or
uninitialized. The plug cannot be used.
‘Check network connectors’
message shows on the monitor
screen
Monitor-Network cable Cable not connected properly to the wall
plate or to the monitor.
Cable or connector defective.
Identification plug on
UPI4NET board
There is no identification plug properly
attached to the monitor.
The identification plug is defective or
uninitialized. The plug cannot be used. See
network service page for details.
‘Network board error’ message
shows on the monitor screen
B-CPU6 board The B-CPU6 board is defective. The board
cannot be used. See network service page
for details.
Other Site View shows no
waveforms
No waveforms are set up
for Monitor-to-Monitor
communication
Run Network Setup to verify current
Monitor-to-Monitor communication setup.