Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
30
Document no. M1137263-04
The different mounts to install the Central Unit F-CU5
(1) Wall Mount Hanger for S/5 F-CU5
(2) Power Adapter Mount for F-CU5 Power Unit N-AC
(3) Adapter Plate for S/5 F-CU5
(4) Table Holder for S/5 F-CU5
(5) Holder for S/5 Command Bar
3.4.2 Connecting to mains
Make sure the cables connecting the different units together are connected and secured.
Connect the power cord to the mains power inlet on the N-AC power unit, and to the wall socket.
WARNING The power cord may only be connected to a three-wire, grounded, hospital
grade receptacle.
3.4.3 Connecting to the Datex-Ohmeda Network
To connect the monitor to the Datex-Ohmeda Network, make sure a Network Board B-UPI4NET
is installed.
Use the Monitor-Network cable to connect the monitor to the network as follows:
1. Make sure that the power to the Central Unit is turned off.
2. Connect the RJ-45 network connector to the X10 connector on CPU board, B-CPU6 and
the Identification Plug to the corresponding connector on the B-UPI4NET.
3. Connect the other RJ-45 connector to the corresponding connector on the wall box.
4. Turn on the power to the Central Unit. Enter the Network service menu:
Monitor Setup -
Install/Service (password 16-4-34) - Service (password 26-23-8) - Frame – Network
5. Make sure that the monitor's network communication has been set according to the used
network software:
Network software S-CNET01 -> DRI Level = 2001
Network software S-CNET02 -> DRI Level = 2001
Network software L-NET03 -> DRI Level = 2003
Network software L-NET05 -> DRI Level = 2005
1
2
4
3
5