Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
58
Document no. M1137263-04
4.4.1 Interconnection
Use the INT-External Device Cable, except for Julian and Cicero EM, in which case use the
INT-Julian/Cicero EM cable.
1. Make sure that the power to both monitors is turned off.
2. Connect the 9 pin D-connector to the Interface Module, E-INT. Tighten the finger screws.
3. Connect the 25 pin D-connector to a corresponding connector on the anesthesia
machine. The connector is labeled 'RS-232-C'. Tighten the finger screws.
4.4.2 Setting communication parameters
Set the communication parameters for Cicero and Cato to RS-232: 9600, e, 8, 1 (MEDIBUS 3.00).
In all cases please refer to the documentation provided by Dräger.
4.4.3 Setting interfacing parameters on the S/5 Anesthesia Monitor or
S/5 Critical Care Monitor
1. Press the Monitor Setup key.
2. Select Install/Service (password 16-4-34).
3. Select Service (password 26-23-8) - Parameters - More...
4. Select Interface and combine a required parameter with the external monitor. Two
letters, al, denote alarm integration. The selection will be automatically stored in
permanent memory.
5. Press the
Normal Screen key.
Table 9 Parameters transferred from Dräger Cicero monitor to S/5
Anesthesia Monitor or S/5 Critical Care Monitor
Selection Waveforms Numerics States
Gases / SSS CO
2
(kPa, mmHg, %) CO
2
Fi & Et (kPa, %)
O
2
Fi & Et (%)
N
2
O Fi & Et (%)
RR (CO
2
) (1/min)
AA Fi & Et (%)
Tidal Volume insp. (1)
Minute Volume exp. (1)
Ppeak (mbar)
Pplat (mbar), PEEP (mbar)
None
SpO
2
Pleth (% full scale) SpO
2
(%)
Pulse rate (1/min)
Probe Off