3.3 Disassembly and reassembly
3.3.1 Command Bar, K-ANEB/ K-ICUB
The Command Bar is disassembled according to the following procedure. Please refer to the
exploded view of the Command Bar.
1. Verify that the monitor is set to standby and Disconnect the Command Bar cable from the
Central Unit.
2. Remove the Command Bar from the display unit.
3. Unscrew the three screws holding the front cover of the Command Bar to the rear cover.
Carefully remove the rear cover.
4. Disconnect the Command Bar cable, the wire set from the ON/STBY switch, the wire set
from the ComWheel (K-ANEB, K-ICUB) and the flat cable from the membrane keyboard
(K-ANEB, K-ICUB).
5. Detach the Command Bar PCB by pressing the two plastic fasteners holding the PCB in
place and simultaneously lifting off the PCB.
6. The Command Bar is reassembled by reversing the disassembly procedure.