3.3.5 Power Unit N-AC
Reassemble the device in reverse order.
When reinstalling make sure the cables connecting the units together are properly connected
and secured.
To remove the CPU mother board
1. Remove the F-CPU bottom cover.
2. Detach the four screws holding the back plate to the
EMC cover.
3. Remove the four screws to detach the CPU mother
board from the back plate.
To reassemble and replace the fan unit
1. Turn the Power Unit upside down.
2. Remove the four TORX screws (T10) to remove
the bottom plate.
3. Detach the fan unit.