Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
12
Document no. M1137272-02
2.3.1 Output connectors, B-DISPX-01
DVI-interface
Pin DIR Level Type Description
1 TX2- O TMDS DIF Differential data to display
2TX2+ O TMDS DIF
32/4_SHIELD G GND Shield for data pairs 2 & 4
4 TX4+ O TMDS DIF Differential data to display
5TX4- O TMDS DIF
6 DDC_CLOCK O OC PU5
7 DDC_DATA I/O OC PU5
8 ANA_VSYNC O CMOS Vertical sync for analog display
9 TX1- O TMDS DIF Differential data to display
10 TX1+ O TMDS DIF
11 1/3_SHIELD G GND Shield for data pairs 1 & 3
12 TX3- O TMDS DIF Differential data to display
13 TX3+ O TMDS DIF
14 +5V_OUT PO +5V +5V supply to display
15 GND G GND
16 HP_DET I +5V HotPlug detection
17 TXD0- O TMDS DIF Differential data to display
18 TXD0+ O TMDS DIF
19 0/5_SHIELD G GND Shield for data pairs 0 & 5
20 TX5- O TMDS DIF Differential data to display
21 TX5+ O TMDS DIF
22 CLK_SHIELD G GND Shield for CLK pair
23 CLK+ O TMDS DIF Differential CLK to display
24 CLK- O TMDS DIF
C1 ANA_RED O ANA Red, Green and Blue for analog display
C2 ANA_GREEN O ANA
C3 ANA_BLUE O ANA
C4 ANA_HSYNC O CMOS Horizontal sync for analog display
C5 ANA_RETURN G GND Return ground for analog interface
18
17 24
C1 C2
C3 C4
C5