Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
44
Document no. M1137266-02
4.12 Tonometry
Service menu for Gastric Tonometry Module.
Service Data
Tonometry: A selection to enter the Tonometry setup
menu.
PressSensCal: A selection to enter the module
pressure sensor calibration menu.
System Test: A selection to enter the module system
test.
CO2 Sensor: A selection to enter the module CO
2
sensor menu.
Pneumatics: A selection to enter the module
pneumatics menu.
AD: A selection to enter the module AD value display.
Record Data: A selection to print out the shown
service data to the device defined in the ...Service -
Record Data menu.
TONO
Statuses (HEX): See APPENDIX A: How to read HEX numbers.
General: Module general status. See Table 2 Module general status for a detailed description of the
message.
Table 2 Module general status
bit 0-5 Not used
bit 6 State error
bit 7 Communication failure
bit 8 Power failure
bit 9 Clock failure
bit 10 EEPROM checksum failure
bit 11 EEPROM writing failure
bit 12 ROM failure
bit 13 RAM failure
bit 14 Test mode
bit 15 Init mode