EasyManua.ls Logo

GE Datex-Ohmeda S/5 - Figure 8 Module Insert; Inserting the Parameter Modules

GE Datex-Ohmeda S/5
453 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System installation
23
Document no. M1137263-04
6. Confirm that the network symbol and ‘Connected to Network’ message are displayed on
the upper part of the screen.
3.3.3 Inserting the parameter modules
1. Ensure that the module is properly orientated (i.e. module release latch facing downward).
2. Align the module insertion guide slot with the frame insertion guide.
3. Push the module into the frame until it clicks.
Figure 8 Module insert
NOTE: The Compact Airway Modules cannot be placed into the third and fourth slot from the
right-hand side of the F-CU8 Central Unit.
NOTE: Use only one Extension Frame F-EXT4 in one F-CU8 Central Unit.
NOTE: Do not use two or more parameter modules with identical functions in the monitor
system. Take special care not to do this when using the Extension Frame, F-EXT4.
Hemodynamic multiparameter modules, E-PRESTN, E-PRETN, E-RESTN, E-PSM, E-PSMP,
M-PRESTN, M-PRETN, M-RESTN, M-ESTPR, M-ETPR, M-ESTR, M-NESTPR, M-NETPR,
M-NESTR, M-NE12STPR, M-NE12TPR, M-NE12STR
Dual Pressure Modules, E-PP/M-PP
Pressure Modules, E-P/ M-P, Pressure Temp Module, E-PT/ M-PT
Cardiac Output Modules, E-COP, E-COPSv, M-COP and M-COPSv
NIBP Modules, M-NIBP and hemodynamic modules w/ N measurement
Airway Modules, E-CO, E-COV, E-COVX, E-CAiOVX, E-CAiOV, E-CAiO, E-miniC, M-C, M-CO,
M-COV, M-CAiO, M-CAiOV, M-CAiOVX, G-O, G-OV, G-AO, G-AiO, G-AOV, G-AiOV, M-MiniC
Interface Module, E-INT / M-INT and Interface Board, B-INT
Oxygen Saturation Modules, E-MASIMO, E-NSATX/ E-NSAT/ M-NSAT and M-OSAT
Tonometry Modules, E-TONO/ M-TONO
NeuroMuscular Transmission Modules, E-NMT/ M-NMT
EEG Modules, E-EEG/ M-EEG
BIS Modules, E-BIS/ M-BIS

Table of Contents

Related product manuals