EasyManua.ls Logo

GE Datex-Ohmeda S/5 - 2 Functional Description; 8-Module Frame, F-CU8

GE Datex-Ohmeda S/5
453 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S/5
TM
8-Module Frame, F-CU8 -12
3
Document no. M1125635-05
2 Functional description
2.1 8-Module Frame, F-CU8
To operate S/5 Anesthesia Monitor or S/5 Critical Care Monitor, the following products should
be installed into the frame
CPU board, w/ main software
UPI4NET board, B-UPI4NET
Optional display controller board(s), B-DISPX (up to two) for secondary and 3rd display
From F-CU8 Rev.12 onwards the CPU board, B-CPU6 with an integrated display and Ethernet
controller and the UPI4NET board, B-UPI4NET are installed into the 8-Module Frame at the
factory.
The frame has two sections. The front part is for the modules. Between the front and rear parts,
there is a module mother board and a CPU mother board. The module mother board connects
modules to the system, and the CPU mother board connects boards together.
Figure 2 Basic structure of S/5 Monitor, an example of possible configuration
Parameter module
Parameter module
Parameter module
Parameter module
Parameter module
CPU Mother Board
Module Mother Board
Display
Command
Board
Optional
display
controller
board
UPI
Board
CPU
Board with
integrated
display
controller
Power Supply
Unit
Module Bus
CPU Bus
Fcu8struct_with_CPU6.vsd