3 Service procedures
3.1 General service information
The field service of the F-CU8 is limited to replacing the faulty printed circuit boards or
mechanical parts. The printed circuit boards should be returned to GE Healthcare for repair.
GE Healthcare is always available for service advice. Please provide the unit serial number, full
type designation, and a detailed description of the fault.
WARNING Only trained personnel with appropriate equipment should perform the
tests and repairs outlined in this section. Unauthorized service may void
warranty of the unit.
3.2 Service check
These instructions include complete procedures for a service check. The service check is
mandatory after any service repair. However, the service check procedures can also be used
for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form (APPENDIX A:) which may be used when performing the
procedures.
The symbol
"in the instructions indicates that the check form contains space to record the
results of the particular procedure.
3.2.1 Recommended tools
3.2.2 Recommended parts
Tool Order No. Notes
Command Bar
E-REC
Hemodynamic Multiparameter Module
Patient Simulator
Multimeter
Screwdriver
Board removal tool or pliers
Antistatic wristband
Part Order No. Notes
Recorder paper
Fan filter 871558