Service Menu
53
Document no. M1137266-02
4.12.5 Pneumatics
Zero valve: A selection to toggle the zero valve on/off.
Press valve: A selection to toggle the pump valve on/
off.
Record Data: A selection to print out the shown
service data to the device defined in the ...Service -
Record Data menu.
TONO
Cathpres; pressure measured from the catheter line
[mbar].
Ambpres; ambient pressure in the room [mbar].
Pumppres; pressure measured from the pumping unit
[mbar].
Pneuma Power; displays whether electricity has been
connected to the pneumatics system or not.
Zero valve; displays whether the zero valve is on or
off.
Pump valve; displays whether the pump valve is on or
off.
Gas removed count; displays how many times gas is removed from the catheter during a
measurement cycle.
Pneumatic errors; detailed information of the performance of different pneumatics system
parts. The meaning of the messages:
‘NO’, the part performed OK during the measurement = ‘this error did not occur’
‘YES’, the part failed during the measurement = ‘this error did occur’
See Table 6 Module pneumatics error status for a detailed description of the messages.