Service Menu
7
Document no. M1137266-02
Table 1 Ethernet service data
Value Usage Notes
Received packets (Statistics
In/Packets)
Total number of received packets since
last cold start.
Transmitted packets (Statistics
Out/Packets)
Total number of transmitted packets
since last cold start.
Received bytes (Statistics In/Bytes) Total number of received bytes since
last cold start.
Transmitted bytes (Statistics
In/Bytes)
Total number of transmitted bytes since
last cold start.
CRC errors (CRC) Number of received packets with
incorrect checksum.
Frame errors (Frame) Number of received packets with
incorrect frame structure.
Refers to physical layer
problems. An erroneous packet
often has both frame and CRC
error.
Transmission errors (Transm.) Number or errors in packet
Transmission.
Internal errors (Intern.) Internal error of the network board. Must always be 0.
Missed packets (Missed) Number of received packets lost due to
overload.
Must always be 0.
FIFO errors (FIFO) Internal error of the network board. Must always be 0.
Overrun errors (Overrun) Practically same as above. Must always be 0.
1.3.3 Ethernet
The Ethernet Status view shows the general
status of the ethernet network communication.
Driver: Ethernet chip name (82559ER)
Cable: Indicates if the ethernet cable is
connected.
EthernetAddr: Monitor’s ethernet address.
Speed: Indicates the current ethernet
communication speed.
The service data related to the ethernet status
view is described in the following table.