Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
6
Document no. M1137266-02
Su
b
net Status
The Subnet status view gives more accurate
information of the different subnet id:s connected. All
four Subnet status menus have a similar structure.
The number of different packets transmitted and
received by the monitor are shown in the columns
below Tx and Rx. The packet types are:
Waveforms: Waveform data
Phys. data: Physiological numerical data
Alarms: Alarms, alarm profiles and alarm limits
Link mgmt: Network management messages
Record K: Record Keeper data
MonToMon: Monitor-to-monitor communication
related data
Printer: Printing data and control messages
File Op.: File operation messages, saving and
loading of cases
Service: Maintenance and service
Modes: User mode data
Indics.: Remote indications sent to monitor
RemoteEv: Remote events
Data server: Packets of the data server (Arrhythmia Workstation)
Packets total: Total number of packets sent/received
Bytes total: Total number of bytes sent/received
T-o Number or resendings
InE Received faulty packets
LenE Erroneous packet length
Dupl Same packet received as a duplicate
1.3.2 Network Config
The DRI Level is for setting the monitor's network
communication. The network communication is
set according to the network software used (e.g.
S-CNET01).
Network software S-CNET01 ->DRI Level = 2001
Network software S-CNET02 ->DRI Level = 2001
Network software L-NET03 ->DRI Level = 2003
Network software L-NET05 ->DRI Level = 2005
Network software L-NET05-02 ->DRI Level = 2008