5-Module Frame, F-CU5, F-CU5P
3
Document no. M1125636-05
2 Functional description
2.1 F-CU5 Central Unit
F-CU5 Central Unit provides places for up to five single-width modules or two double-width
modules (plus one slot for one single-width module). It is also called 5-Module Frame.
F-CU5P with E-PSM(P) support provides places for up to five single-width modules or two
double-width modules (plus one slot for one single-width module) and a connector plate for
Patient Side Module E-PSM or E-PSMP. It is also called 5-Module Frame.
The F-CU5(P) Central Unit option includes three components:
• 5-Module Frame, F-CU5(P)
• Central Processor Unit, F-CPU
• Power Unit, N-AC
The units are connected together with appropriate cables:
• F-CPU Power Cable (0.4m/16 inch or 2.7m/9ft)
• F-CPU Signal Cable (0.3m/12inch, 3m/10ft or 10m/33ft)
NOTE: You can connect two F-CU5(P) 5-Module Frames to one F-CPU. If you do so, you cannot
at the same time use a display that uses B-DISPX as power supply (for example 12" LCD
display).
Figure 2 5-Module Frame F-CU5P w/ modules
To operate S/5 Anesthesia Monitor or S/5 Critical Care Monitor with 5-Module Frame, F-CU5(P),
the following products should be installed:
• CPU board, B-CPU6 w/ L-ANE07(A) or L-ICU07(A)
• UPI4NET board, B-UPI4NET revision 01 or later
• Optional display controller board, B-DISPX
The CPU board, B-CPU6, the UPI4NET board, B-UPI4NET and the optional display controller
board, B-DISPX are included and installed in the Central Unit, F-CPU at the factory.
NOTE: The B-INT cannot be used in the F-CPU.