Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
68
Document no. M1137263-04
5.3.4 5-Module Central Unit, F-CU5/ 8-Module Central Unit, F-CU8
1. Check that all the fans in the N-AC, F-CPU and F-CU5(P) are / the frame fan in F-CU8 is
running.
2. Check that the clock on the screen shows correct time. Readjust the time and date, if
necessary.
"
5.3.5 Extension Frame, F-EXT4
1. If the F-EXT4 contains a fan, check that the fan is running.
2. Check that the modules in the F-EXT4 are recognized i.e. the required parameter
information is shown on the monitor screen.
"
5.3.6 Compact Airway Module, E-CXXXXX
Wait until the message ‘Calibrating gas sensor’ disappears from the screen.
1. Check that the fan is running.
2. If the module contains membrane keys on the front panel, check the function of each of
the membrane keys.
"
Compact Airway Modules with the Patient Spirometry option
3. Connect a clean spirometry tube and D-lite to the module. Connect the sampling line.
Breathe through the wider side of the D-lite.
Check that the flow waveform moves downwards when you breathe in and upwards
when you breathe out.
"
For all Compact Airway Modules
4. Block the tip of the sampling line with your finger and check that the message ‘Sample
line blocked’ appears on the monitor screen within 30 seconds.
5. Detach the D-fend and check that the message ‘Check D-fend’ appears on the monitor
screen within 30 seconds.
6. Breathe to the sampling line briefly. Check that the CO2 information is updated on the
screen.
"