S/5 Extension Frame, F-EXT, Extension Module, E-EXT
7
Document no. M1125641-03
3 Service procedures
The field service of the Extension Frame, F-EXT4, is limited to replacing faulty circuit boards or
mechanical parts. The circuit boards should be returned to GE Healthcare for repair.
GE Healthcare is always available for service advice. Please provide the unit serial number, full
type designation, and a detailed description of the fault.
WARNING Only trained personnel with appropriate equipment should perform the
tests and repairs outlined in this section. Unauthorized service may void
warranty of the unit.
3.1 Service check
These instructions include complete procedures for a service check. The service check is
mandatory after any service repair. However, the service check procedures can also be used
for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form (“APPENDIX A:”) which may be used when performing the
procedures.
The symbol
"in the instructions indicates that the check form contains space to record the
results of the particular procedure
3.1.1 Recommended tools
3.1.2 Visual inspection
•
Disconnect the Extension Module, E-EXT from the host monitor, if connected.
• Disconnect the extension module cable from the Extension Frame, F-EXT4 rear panel.
Extension Module, E-EXT
•
Detach the module box by removing the two screws from the back of the module. Be
careful with the loose latch and spring pin for locking.
1. Internal parts
Check that:
− all screws are tightened properly
− there are no loose objects inside the module
"
Tool Order No. Notes
Hemodynamic module W/NIBP
Adult NIBP cuff & hose
Gas interface cable 884299
Multimeter
Screwdriver